Computer
System Validation
Basic
Concepts of Computer System Validation in
Pharmaceutical
Industry
What
is Computer System?
Computerized systems are the
system which used extensively to capture, analyze, store data, report
information and to control equipment and processes in life sciences industry.
Software
Validation
Provision of objective evidence that software
specifications conform to user needs and intended uses and that the particular
requirements implemented through software can be consistently fulfilled
Computer
System Validation Cycle
1.
Validation
Plan
2.
User
Requirements Specifications
3.
Functional
Specifications
4.
Design
Specifications
5.
System
Developments
6.
Installation
Qualification (IQ)
7.
Operational
Qualification (OQ)
8.
Performance
Qualification (PQ)
1.
Validation Plan
The
plan typically describes work to be done, resources required, methods to be
used, configuration management and quality assurance procedures to be followed
thought activity.
1.
Scope
2.
Approach
3.
Roles
& Responsibilities
4.
Acceptance
Criteria
2.
User Requirement Specifications
The
User Requirements Specification describes the business needs for what users
require from the system. URS are written by the system owner and
end-users, with input received from Quality Assurance.
1.
Business
&User Needs from the system.
2.
Intended
Use of the system
3.
Critical
Constraints
3.
Functional
Specification
Functional
specification that specifies the functions that a system or software must
perform which predefined in URS.
1.
Verifies
specified function
2.
What
data the software should capture
3. Logic,
calculations which used in application.
4.
Design
Specification
Design
Specifications describe how a system performs the requirements outlined in the
Functional Requirements & User Requirement Specification.
1.
Front
End Design
2.
Process
design
3.
Database
design
4.
Security
design
5.
Interface
design
6.
Network
requirements
5.
System
development
Systems development is
the process of defining, designing, testing, and implementing a new software
application or program.
6.
Installation
Qualification
The
Installation Qualification Protocol verifies the proper installation and
configuration of a System. This can include ensuring that necessary application
files have been loaded, equipment has been installed, the necessary procedures
have been approved, or the appropriate personnel have been trained.
7.
Operational
Qualification
The Operational
Qualification Protocol is a collection of test cases used to verify the proper
functioning of a system. The operational qualification test requirements are
defined in the Functional Requirements specification. Operational Qualification
is usually performed before the system is released for use.
Ø
Confirmation
that all functionality is present
Ø
Confirmation
that all feature are working as specified
8.
Performance
Qualification
Performance
Qualifications are a collection of test cases used to verify that a system
performs as expected under simulated real-world conditions. The performance
qualification tests requirements defined in the User Requirements Specification (or possibly
the Functional
Requirements Specification).
Terminology
I.
VALIDATION
Confirmation by examination and provision
of objective evidence that software specifications conform to user needs and
intended uses, and that the particular requirements implemented through
software can be consistently fulfilled.
II.
VERIFICATION
Software verification looks for
consistency, completeness, and correctness of the software and its supporting
documentation, as it is being developed, and provides support for a subsequent
conclusion that software is validated.
III.
QUALIFICATION
Formal
testing to demonstrate that the software meets its specified requirements.
General Concepts
I.
Validation scope
can be limited to the features that will be used by the regulated company.
Explanation
A device manufacturer who chooses
not to use all the vendor supplied capabilities of the software only needs to
validate those functions that will be used and for which the device
manufacturer is dependent upon the software results as part of production or
the quality system
II.
Validation must
be specific to the regulated company’s planned and documented use of the
application
Explanation
The acceptance of vendor supplied
validation data in isolation of system configuration and intended use is not
acceptable. In isolation from the intended process or end user IT
infrastructure, vendor testing is likely to be limited to functional
verification only, and may not fulfill the requirements for performance qualification.
III.
Vendor
documentation can be used as the starting point for validation
Explanation
If the vendor can provide information
about their system requirements, software requirements, validation process, and
the results of their validation, the medical device manufacturer can use that
information as a beginning point for their required validation documentation. FDA,
General Principles
IV.
The regulated
company needs to audit the vendors of critical applications and services –
depending on risk
Explanation
The audit should demonstrate that
the vendor’s procedures for and results of the verification and validation
activities performed for the OTS software are appropriate and sufficient for
the safety and effectiveness requirements …
V.
Service Level
Agreement (SLA) Formal agreements are required to document responsibilities
Explanation
When third parties
(e.g.suppliers, service providers) are used e.g. to provide, install, configure, integrate, validate, maintain
(e.g. via remote access), modify or retain a computerised system or related
service or for data processing, formal agreements must exist between the
manufacturer and any third parties, and these agreements should include clear
statements of the responsibilities of the third party.
Very good knowledge improvement
ReplyDeleteExcellent knowledge building content. Keep it up.
ReplyDelete